Portex Spinal Tray with 25G Whitacre Needle. Kit contains Bupivacaine in Dextrose, Epinephrine,Lidocaine, Lidocaine in Dextrose and Tetracaine Catalog Number: 15545-20
FDA Recall Z-0166-04 · Posted 2003-11-27
Advertisement
Recalling FirmPortex Inc
Recall StatusTerminated
Product CodeCAZ
Date Initiated2003-10-10
Date Terminated2006-05-02
Product Quantity2600 units
Reason for Recall
Spinal trays labeled as containing a Whitacre 25g x 3.5' needle actually contain a Quincke 25g x 3.5' needle.
Firm Action
Portex notified consignees on 10/10/03 by letter issued by Federal Express. Distributors were asked to supply customer list to Portex or to notify their accounts with copy of the recall letter provided.
Root Cause
Other
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.