Siemens brand 550 TxT Patient Table, Part No. 7346534, Distributed by and/or Manufactured by Siemens AG, Medical Solutions, Germany
FDA Recall Z-0166-2011 · Posted 2010-10-28
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeJAI
Date Initiated2010-03-31
Date Terminated2012-03-01
Product Quantity326 units
Reason for Recall
An increase in voltage may cause a short circuit of a power transistor that may cause a smoldering and smoking effect of the electrical wiring and also an unexpected lowering of the Table. Both issues present the potential for injury to a patient or the operator.
Firm Action
Siemens sent a Customer Information Letter on March 31, 2010, to all consignees. The letter identified the product, the problem, and the action that should be taken. Siemens implemented minor modifications to address the issues. No further action was required for the continual operation of the 550TxT Table.
For questions regarding this recall call 925-602-8083.
Root Cause
Other
Distribution Pattern
Worldwide Distribution including USA