Tubing, Peristaltic Head Part Number 7-35009685-01 for use in clinical chemistry systems.

FDA Recall Z-0166-2018 · Posted None

Recall StatusCompleted
Product CodeJJE
Date Initiated2017-09-29
Product Quantity44 units

Reason for Recall

The Peristaltic Head tubing placed on the ARCHITECT c4000, c8000, and c16000 has the possibility to leak due to a manufacturing issue associated with the tubing connector.

Firm Action

All worldwide ARCHITECT Clinical Chemistry System customers with impacted instruments will receive the Product Correction Letter dated, 9/29/17. A customer reply is requested from all affected consignees.

Root Cause

Component design/selection

Distribution Pattern

Worldwide distribution to US, Canada, France, and Germany.