ROSA Brain 3.0.0 Usage: The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic surgery may be appropriate.

FDA Recall Z-0167-2018 · Posted None

Recalling FirmZimmer Biomet, Inc.
Recall StatusTerminated
Product CodeHAW
Date Initiated2016-07-20
Date Terminated2020-06-16
Product Quantity18 (14 US and 4 OUS)

Reason for Recall

Communication errors between ROSANNA BRAIN software, MARIO software and the Stubli CS8C controller.

Firm Action

This is a retrospective report of a correction initiated on 20-Jul-2016. Customers were informed onsite by field Service Technicians of the planned correction. The software issue described was corrected in the new software version ROSA Brain 3.0.0.20. Field Service Technicians were deployed to the customers locations to perform the system upgrade. .

Root Cause

Software design

Distribution Pattern

Worldwide Distribution - US Nationwide in the states of OH, FL, MA, MN, DC, NC, CA, NY, PA and countries of Australia and France