Terminated

Byrd Screw, 2.0X6MM SD, Model Number: 01-0264 Product labeling reads in part as follows: REF: 01-0264, QTY,UOM 00001, BYRD SCREW, 2 OX6MM SELF-DRILLING, 12CM, (4-3/4) (L), STAINLESS STEEL, LOT 999999, (01) 00841036007157(10)999999(13)090201(30)13, 2009-02, ***NON-STERILE PRODUCT***DD 14 03 1, BIOMET MICROFIXATION, INC., 150 TRADEPORT DRIVE, JACKSONVILLE, FL 32218 USA Instrument used to pull up the broken bone, so it is flush and can be plated fixated. The Byrd screw is then removed.

FDA Recall Z-0169-2010 · Posted 2009-11-10

Recall StatusTerminated
Product CodeLXH
Date Initiated2009-02-19
Date Terminated2012-05-03
Product Quantity22

Reason for Recall

Byrd Screw, Self Drilling, may break during usage. BIOMET has determined that the parts were not correctly processed by the vendor , i.e. that the parts were not hardened in the final processing and that lead to weaker than acceptable material.

Firm Action

Biomet plans to send letters to customers on February 19, 2009. Customers are told to discontinue use of the item and return it to Biomet Microfixation. The firm plans to follow up with an e-mail to their customers. Questions pertaining to this issue, should be directed to Christy Cain at 1.800.874.7711 or 904.741.4400.

Root Cause

Nonconforming Material/Component

Distribution Pattern

AZ, CA, GA, IL, IN, NM, NY, TX, VA, Puerto Rico and Canada.