Signa MFO/I Magnetic Resonance System, Model Numbers : 2210339, 2262700, 2276937, 2282853, 2282860, 2282880, 2352129, 2388781, 5105123, 5113339, 2138300-32, 2189500-3, 2198999-4, 2259620-4, 2262700-2, 2269512-2, 2286849-2, 2352129-2, 2377062-5, 2377062-8, Manufactured by GE Hangwei Medical Systems Co., LTD, Beijing PR, China

FDA Recall Z-0170-2008 · Posted 2007-11-07

Recalling FirmGE Healthcare
Recall StatusTerminated
Product CodeLNH
Date Initiated2006-12-31
Date Terminated2007-12-11
Product Quantity163

Reason for Recall

Incorrect patient positioning; when moving the patient into the bore may lead to wrist and hand injury.

Firm Action

Field Modification Instruction : FMI 67631 (DEC 2006) was communicated directly to the GE Healthcare service engineers who provide the modification at the customer site. Corrections were performed in DEC 2006

Root Cause

Labeling design

Distribution Pattern

Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, FL, ID, IL, IN, KS, KY, LA, MD, MO, MS, MT, NJ, NY, OH, OK, PA, SD, VA, and WA, and countries of United Arab Emirats, Argentina, Belgium, Brazil China, Ivery Cost, Germany, Spain, France, United Kingdom, Hungary, Italy, Japan, Korea, Mexico, Oman, Panama, Puerto Rico, Russia, and Turkey

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.