Eon Rechargeable Implantable Pulse Generator (IPG) System
FDA Recall Z-0170-2010 · Posted 2009-11-10
Advertisement
Recalling FirmAdvanced Neuromodulation Systems, Inc
Recall StatusTerminated
Product CodeLGW
Date Initiated2006-12-22
Date Terminated2010-07-23
Product Quantity24,750 units total
Reason for Recall
Product's external charging system on occasion was failing to communicate with the Implantable Pulse Generator preventing re-charging of the pain relief system.
Firm Action
12/22/2006, Marketing Update 5015 was issued to field sales representatives providing information on improving communication between the Internal Pulse Generator and its charger. May 15, 2007, Marketing Update 5022 was issued to field sales representatives to reiterate avoiding contact between the charger antenna and conductive materials and place the preferred side of the charging antenna face-down over the implant. Troubleshooting Guide was created in January 2008 for field sales representatives and technical services departments regarding isolation of the communication issue to the external devices instead of the implanted component. March 25, 2008, the Clinician's Manual was updated adding a recommendation for a minimum depth of pocket for device implantation.
Root Cause
Device Design
Distribution Pattern
Worldwide
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.