Recall StatusOpen, Classified
Date Initiated2022-09-21
Product Quantity2018
Reason for Recall
Detachment controller, of an aneurysm embolization system, has an out of specification tunnel opening dimension, which may prevent the implant device pusher from being fully inserted into the controller, which may result in the inability of the controller to properly detach the implant device, which could cause delays and excessive manipulation that may cause the implant to move and damage vessels
Firm Action
On 09/21/2022, recall notices were mailed to customers who were asked to do the following:
1) Identify, quarantine, and return all impacted devices.
2) Complete and return Customer Acknowledgment and Device Reconciliation form
the recall instructions provided in the letter and have shared this notification with all device users within my facility and network to ensure they are aware of this recall. This recall notice should also be shared with any organization where the potentially affected devices have been transferred.
Complete and return the acknowledgement and receipt form.
Customers with additional questions can call the recalling firm at 949-501-0355, Monday - Friday PST 8:00 - 5:00 pm or email
[email protected]
Root Cause
Process control
Distribution Pattern
Worldwide - US Nationwide distribution in the states of MN, PA, NY, MI, FL, VA, TX, OH, CA, NE, MN, NC, LA, IL, TN, IN, UT, KS and the countries of Japan, China, Argentina, Australia, South Korea, Italy, UK.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.