Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Surgical Unit (Catalog Number X9007696), and Cockpit, ISU Configuration (Catalog Number X9007707), Software (SW) version 2.7.4; robotic surgical system
FDA Recall Z-0170-2024 · Posted 2023-10-30
Advertisement
Recalling FirmAsensus Surgical Inc
Recall StatusTerminated
Product CodeNAY
Date Initiated2023-09-15
Date Terminated2024-09-26
Product Quantity21 units
Reason for Recall
Asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparoscope Instrument Actuator (LIA) where the LIA rotated continuously in one direction after the Surgeon removed the engagement of teleoperation on the Senhance System. There was no other uncontrolled arm motion observed.
Firm Action
Asensus Surgical issued an URGENT MEDICAL DEVICE RECALL to its consignees on 09/15/2023 by mail. The notice explained the problem, risk to health, and requested users to cease use of the device until software version 2.7.5 is installed.
Root Cause
Software design
Distribution Pattern
US: NJ, LA, FL, MN, IL, and Germany, Japan, Lithuania
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.