Plum 360 Infusion System With MedNet/ Smart Card Plug And Play Module, Model # 30010. Infusion pump.

FDA Recall Z-0171-2020 · Posted 2019-10-22

Recalling FirmICU Medical Inc
Recall StatusTerminated
Product CodeFRN
Date Initiated2018-08-01
Date Terminated2021-04-27
Product Quantity47,152 devices

Reason for Recall

A software malfunction of the infuser when used with a MedNet Custom Drug Library (CDL). When a new CDL is downloaded and installed on the infuser the infuser can become inoperable and show a blank screen on the User Interface.

Firm Action

On 8/1/2018 a "Urgent Medical Device Correction" letter was mailed to consignees via traceable mailing. The medical device correction letter informed consignees of the following: 1. To reduce the likelihood of the issue occurring at the bedside, you can consider not utilizing the option to "Force Pump to Accept Drug Library During Power Down" when finalizing the CDL with MedNet. In doing so, when powering down the infuser the clinician will have the option to install the CDL or not, based on the clinical situation. 2. ICU Medical will release a software update to correct the issue and will contact you when the update is available. 3. . Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular 4.Please inform potential users in your organization of this notification, complete the attached response form and return it to the fax number (1-888-628-6849 or e-mail address ([email protected]) on the form, even if you do not have the affected product. 5. For further inquiries, please contact ICU Medical using the information provided below: Global Complaint Management - 1-844-654-7780 or [email protected] Technical Support Center - 1-800-241-4002, option 4 or [email protected]

Root Cause

Software change control

Distribution Pattern

US nationwide distribution, and Canada