Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Medline brand medical procedure convenience kits, labeled as: a) UNIVERSAL DRAPE PACK, REF DYNJ44114; b) DRAPE PACK, REF DYNJ82663B

FDA Recall Z-0171-2025 · Posted 2024-10-23

Recall StatusOpen, Classified
Product CodeKKX
Date Initiated2024-09-04
Product Quantity194 units

Reason for Recall

It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.

Firm Action

Medline issued a MEDICAL DEVICE RECALL notice to its consignees on 09/05/2024 via email and USPS. The notice explained the issue, hazard, and requested the destruction of the affected product. Distributors were directed to notify their consignees.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Panama.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑