Advanced D-Dimer used with the Sysmex CA-1500, CA-7000, CA-500 Systems when using the AKIMA method.

FDA Recall Z-0172-06 · Posted 2005-11-18

Recalling FirmDade Behring , Inc.
Recall StatusTerminated
Product CodeDAP
Date Initiated2005-07-29
Date Terminated2006-03-01
Product Quantity2987 units

Reason for Recall

Reports of false negative results on Sysmex Coagulation System.

Firm Action

On 7/29/05, the recalling firm issued a recall letter to their accounts informing them of the probelm. The letter also provided instruction on how to prevent the probelm from occurring - change from the AKIMA method to the Point ot Point method.

Root Cause

Other

Distribution Pattern

The product was shipped to medical facilities nationwide.