Terminated

VITROS VANC (vancoymycin) Reagent, REF #680 1709. Firm on label: Ortho-Clinical Diagnostics, Inc., Rochester, NY 14626

FDA Recall Z-0173-06 · Posted 2005-11-18

Recall StatusTerminated
Product CodeLEH
Date Initiated2005-10-12
Date Terminated2008-09-22
Product Quantity784 boxes x 6 packs x 50 tests

Reason for Recall

Negatively biased results can be observed when the reagent pack has less than five tests remaining and has been on the analyzer greater than 2 days.

Firm Action

Letters dated 10/12/2005 issued to end users in the U.S. Customers advised they may continue to use the VANC Reagent if their laboratory processes QC fluids according to protocol. A return reply form is included. Foreign affiliates were notified via e-mail on 10/12/2005.

Root Cause

Other

Distribution Pattern

Product shipped to medical facilities (end-users) in the U.S. and to 9 International J&J affiliates in Australia, Canada, Caribbean, Chile, France, Philippines, Portugal, Taiwan, UK/Northern Europe, Israel, Saudi Arabia and Venezuela.