Baxter System 1000 Tina Single Patient Hemodialysis Delivery System; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085; product code S1000L3P. Hemodialysis and Continuous Renal Replacement Therapy devices for acute and chronic hemodialysis
FDA Recall Z-0173-2012 · Posted 2011-11-07
Advertisement
Recalling FirmBaxter Healthcare Renal Div
Recall StatusTerminated
Product CodeKDI
Date Initiated2011-09-19
Date Terminated2011-11-07
Product Quantity1 unit
Reason for Recall
The Tina System was released to the customer with OOS values for the Pressure Displays Accuracy and Leakage Rate Test.
Firm Action
Baxter Healthcare Corporation sent a service technician to the lone customer site on September 19, 2011 to correct the affected product. The technician executed all tests as established in the Service Manual, with acceptable results.
Root Cause
Employee error
Distribution Pattern
Distributed in Puerto Rico only.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.