Emerald Diluent Tubing, List number 09H50-01, accompanying CELL-DYN Emerald instrument, List number 09H39-01. The firm name on the label is Abbott Laboratories, Abbott Park, IL.

FDA Recall Z-0173-2020 · Posted None

Recalling FirmAbbott Laboratories
Recall StatusTerminated
Product CodeGKZ
Date Initiated2019-05-09
Date Terminated2020-10-07
Product Quantity22 analyzers with tubing

Reason for Recall

The diluent tubing used on the instrument appears cloudy or to have a powder/film on the inner and outer surfaces which may result in falsely elevated platelet results.

Firm Action

The recalling firm issued letters on 5/9/2019 to the consignees who had the installed instruments with the affected serial numbers. The letters were issued via FedEx priority overnight express.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Distribution was made to GA, LA, MA, NC, NY, SC, and TX. There was no military/government distribution. Foreign distribution was made to Australia, Germany, Greece, Honduras, Indonesia, Italy, and Serbia.