Magnetic Resonance Imaging. Brand names: Concerto, Concerto Upgrade, Syngo Harmony, Syngo Harmony 087S, Syngo Harmony Upgrade, Syngo Sonata, Syngo Sonata Upgrade, Syngo Symphony, Syngo Symphony Q, and Syngo Symphony Upgrade.
FDA Recall Z-0175-03 · Posted 2003-12-03
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeLNH
Date Initiated2003-10-16
Date Terminated2004-08-25
Product Quantity118 units
Reason for Recall
Equipment SAR monitor does not accept new patient data, incorrect RF levels may be administered to the patient.
Firm Action
The recalling firm field representatives are visiting each location to correct the problem through a software upgrade.
Root Cause
Other
Distribution Pattern
The products were shipped nationwide to hospitals and medical imagining centers.