Terminated

Patient Data Manager, Model/Catalogue Numbers: 25100-09 - PATIENT DATA MANAGER 2.1.1LICENSE ONLY 25100-10 - PATIENT DATA MANAGER 2.2.0 LICENSE ONLY 25100-11 - PATIENT DATA MANAGER 2.2.1 LICENSE ONLY 25100-12 - PATIENT DATA MANAGER 2.3 INSTALLER 25100-12A - PATIENT DATA MANAGER 2.3.1 INSTALLER 25100-13 - PATIENT DATA MANAGER 2.3 LICENSE 25100-12B - PATIENT DATA MANAGER 2.4.0 INSTALLER 25100-13A - PATIENT DATA MANAGER 2.4 LICENSE 25100-12C - PATIENT DATA MANAGER 2.5.0 INSTALLER 25100-12D - PATIENT DATA MANAGER 2.5.1 INSTALLER 25100-13B - PATIENT DATA MANAGER 2.5 LICENSE used in combination with iPlan CMF/Cranial/ENT/Spine 3.0.5, 3.0.6. Product Usage: Patient Data Manager is intended to load patient data from and export patient data to connected network locations (like file shares and PACS) and removable media. Furthermore, the software is intended to start other applications (like Brainlab planning and navigation applications) and to route applications and connected video sources to displays attached to the system that are running Patient Data Manager. The software is also intended to connect two Brainlab systems to each other via network.

FDA Recall Z-0176-2020 · Posted None

Recalling FirmBrainlab AG
Recall StatusTerminated
Product CodeLMA
Date Initiated2019-09-25
Date Terminated2024-03-25
Product Quantity1230 units

Reason for Recall

The Brainlab Patient Data Manager software (PDM) in combination with the Brainlab iPlan CMF/Cranial/ENT/Spine software 2.1.1, 2.2.0, 2.2.1, 2.3.0, 2.3.1, 2.4.0, 2.5.0, 2.5.1 (if used in combination with iPlan CMF/Cranial/ENT/Spine 3.0.5, 3.0.6) that may lead to shifts applied to segmentation objects in iPlan being not reflected when exported into DICOM under specific circumstances.

Firm Action

As of September 25, 2019 for new installations or updates/upgrades to the affected SW version, the Field Safety Notice / Product Notification letter is hand delivered to the customer during installation (before clinical use) by the according Brainlab Service representative All other consignees were notified by email on 10/04/2019. The letter details how the even occurs and provides instruction on how to avoid the issue until a software revision is disseminated.

Root Cause

Software design

Distribution Pattern

worldwide