Terminated

Passport 2 with View 12 ECG Analysis Module. It is a transportable, multi-parameter physiological monitor designed to monitor and display the following physiological parameters: ECG, Heart Rate derived from selected sources (Sp)2, ECG, IBP and NIBP), SpO2 level, ST Segment, Arrhythmia, Blood pressure (both invasive and non-invasive), Respiration rate (dirived from ECG or CO2), inspired or expired CO2, temperature , and gases (i.e. five anesthetsia gases, O2, N2O, and CO2)

FDA Recall Z-0177-06 · Posted 2005-11-23

Recalling FirmDatascope Corp
Recall StatusTerminated
Product CodeDRT
Date Initiated2005-09-07
Date Terminated2007-06-06
Product Quantity979 monitors

Reason for Recall

Software anomalies which include shut down while printing, invalid diastolic blood pressure display, and incomplete and inaccurate interpretation reports of 12-Lead ECG data.

Firm Action

Datascope sent out a Field Correction notice on 9/7/2005 to all Hospital administrators who have the monitors. This was sent out via certified mail, return receipt requested. A Datascope Service Rep will conduct and document the corrective action during a visit to the customer.

Root Cause

Other

Distribution Pattern

The monitors have been distributed domestically and internationally to hospitals and clinics. There are 10 Passport 2 monitors distributed to Naval Hospital, Camp Pendleton, CA. There are 12 Spectrum monitors distributed to two US Govt facilities: VA Medical Center, Hines, IL and VA Medical Center, Syracuse, NY.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →