Terminated

Rosa Spine 1.0.2 Stereotaxic Instrument Computer-Assisted Surgical Device The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by surgeons to guide standard neurosurgical instruments during spine surgery. Guidance is based on an intra-operative plan developed with three-dimensional imaging software provided that the required fiducial markers and rigid patient anatomy can be identified on 3D CT scans. The device is indicated for the placement of pedicle screws in lumber vertebrae with a posterior approach.

FDA Recall Z-0177-2018 · Posted None

Recalling FirmZimmer Biomet, Inc.
Recall StatusTerminated
Product CodeOLO
Date Initiated2015-10-29
Date Terminated2020-06-16
Product Quantity2

Reason for Recall

A design change was initiated to update ROSA Spine 1.0.2 to version ROSA Spine 1.0.2.16 to resolve software bugs and improve usability and stability of the ROSA Spine device.

Firm Action

Zimmer Biomet issued a retrospective report of a correction taken on 29-Oct-2015. The software issue described was corrected in the new software version 1.0.2.16. There were 2 copies of the affected software in distribution. Field Service Technicians were deployed to the consignee locations to perform the system upgrade. 100% of affected devices were updated with the software solution. No further action is required.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution including France and Germany