Michigan Instruments Thumper Mechanical CPR Device; Model 1007, Part number 15000.

FDA Recall Z-0178-2007 · Posted 2006-11-14

Recall StatusTerminated
Product CodeDRM
Date Initiated2006-09-14
Date Terminated2006-12-06
Product Quantity639

Reason for Recall

Failure to initiate compressions when turned on.

Firm Action

The Japanese distributor is instructed by letter dated 9/14/06 to drill a small hole in the male checked connector at the base of the column in order to allow rapid bleed off of the pressure and to give customers a revised operators manual.

Root Cause

Other

Distribution Pattern

Japan.