Michigan Instruments Thumper Mechanical CPR Device; Model 1007, Part number 15000.
FDA Recall Z-0178-2007 · Posted 2006-11-14
Advertisement
Recalling FirmMichigan Instruments, Inc.
Recall StatusTerminated
Product CodeDRM
Date Initiated2006-09-14
Date Terminated2006-12-06
Product Quantity639
Reason for Recall
Failure to initiate compressions when turned on.
Firm Action
The Japanese distributor is instructed by letter dated 9/14/06 to drill a small hole in the male checked connector at the base of the column in order to allow rapid bleed off of the pressure and to give customers a revised operators manual.
Root Cause
Other
Distribution Pattern
Japan.