Terminated

Terumo Sarns brand Soft-Flow Aortic Cannula with flange, angled, wire-reinforced, with Luer, 6 mm (18 Fr) OD with 3/8 inch connector, 14 inches (35 cm) long; catalog number 7080.

FDA Recall Z-0181-06 · Posted 2005-11-23

Recall StatusTerminated
Product CodeDWF
Date Initiated2005-10-19
Date Terminated2006-03-10
Product Quantity48

Reason for Recall

The product is labeled as sterile, but the firm has determined that the suture flange may tear through the front of the package.

Firm Action

Consignees were notified via telephone, and a recall letter dated 10/19/05 sent to each consignee, beginning on 10/18/05. International affiliates were notified via email.

Root Cause

Other

Distribution Pattern

California, Indiana, Missouri, Brazil and Germany.