Open, Classified

NaturaLyte Liquid Acid concentrate, Part Number 08-0231-4, four (4) one-gallon bottles per bottle case. For dialysis.

FDA Recall Z-0181-2023 · Posted 2022-11-04

Recall StatusOpen, Classified
Product CodeKPO
Date Initiated2022-10-04
Product Quantity17700

Reason for Recall

The "Nominal Chemical Composition" section of the label incorrectly lists 5.62g/L KCL (potassium chloride). 5.62g/L actually corresponds to the CaCl (calcium chloride) content. The product remains potassium free.

Firm Action

On October 4, 2022, the firm notified affected customers via Important Medical Device Information letters. Customers were reassured that the product was manufactured according to specification and remains potassium-free. Customers may continue to use the product. Please contact FMCRTG Customer Service at 1-800-323-5188 if you have further questions.

Root Cause

Error in labeling

Distribution Pattern

Domestic distribution to AL AZ CA DC FL GA HI IL IN MD MI MO MS NE NJ NY OH PA SD TX VA No foreign distribution.