Ronci Surgical Universal Imaging Extension used with the Ronci Pain Management Table System Catalog Number: R-934-25A

FDA Recall Z-0182-04 · Posted 2003-12-10

Recalling FirmThe OR Group
Recall StatusTerminated
Product CodeFPS
Date Initiated2003-11-06
Date Terminated2005-07-25
Product Quantity17 units

Reason for Recall

Imaging table with extension may fail if the table leg is incorrectly positioned and the load exceeds the maximum weight of 400 lb

Firm Action

The OR Group forwarded by 2nd day mail on 11/6/03 the repair kit and instructions to the accounts.

Root Cause

Other

Distribution Pattern

Nationwide. CA, FL, GA, IA MI, MT, NH, OH, NJ, NY, SC, UT, TX, VA