Combiset Blood Line with BVM for Hemodialysis use Catalog Number: 03-2965-9. Intended for use as the extracorporeal blood circuit during Hemodialysis and is indicated for use with conventional and high flux negative pressure hemodialyzer equipment.

FDA Recall Z-0185-2014 · Posted 2013-11-08

Recall StatusTerminated
Product CodeONW
Date Initiated2013-08-29
Date Terminated2021-02-11
Product Quantity24,202 cases

Reason for Recall

Updated Instructions for Use: Bloodline connections not tightened per instructions in the operator's manual may disconnect and lead to blood loss

Firm Action

Fresenius issued a Field Notification Letter dated 8/29/13 to Patient Clinics/ associated at home patients and provides additional detail instructions to ensure the connection to the transducer protector bloodline female luer interface is secure. This field advisory notice sent by Certified Mail, Return Receipt Requested. Accounts requested to return both the receipt and the completed reply form (enclosed) to confirm receipt of the notice. If you have any questions regarding this field advisory, please contact FMCNA Customer Service Team at 1-800-323-5188. Customer Service will be available Monday-Friday 7am-6pm CST.

Root Cause

Labeling design

Distribution Pattern

Worldwide Distribution-USA (nationwide) and the country of Canada.