CyPass System 241, Intraocular Pressure Lowering Implant, CYPASS MICRO-STENT, CYPASS APPLIER, REF: 8065754001. For ophthalmic use.

FDA Recall Z-0186-2019 · Posted 2018-10-23

Recalling FirmAlcon Research, LTD.
Recall StatusTerminated
Product CodeOGO
Date Initiated2018-08-28
Date Terminated2020-12-18
Product Quantity14264 units

Reason for Recall

Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to the group who underwent cataract surgery alone.

Firm Action

The firm initiated their market withdrawal by letter on 08/28/2018. The firm requested surgeons immediately cease implantation of the device and return any unused product.

Root Cause

Device Design

Distribution Pattern

Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, France, Finland, Germany, Hong Kong, Iceland, Italy, Mexico, Peru, Singapore, Spain, Sweden, Switzerland, United Kingdom, and Uruguay.