Maquet Alphastar 1132.01AX/BX, 1132.02AX/BX, and 1132.03AX/BX Universal Operating Table

FDA Recall Z-0188-2009 · Posted 2008-11-05

Recalling FirmMAQUET Inc.
Recall StatusTerminated
Product CodeFQO
Date Initiated2008-09-16
Date Terminated2010-01-12
Product Quantity141 in US

Reason for Recall

Table tilts: The Alphastar Surgical Table piston rod head may break and cause a full inclination of the table top during Trendelenberg positioning if there is insufficient lubrication.

Firm Action

Maquet sent Recall Notification letters to all customers on September 16, 2008 by Federal Express. Letters inform customers of the potential problem with the Alphastar tables, a description of the failure and hazards involved. Hospitals are asked to confirm whether yearly maintenance by an authorized technician have been performed. If preventive maintenance has not been performed, customers are requested to contact their Maquet Service rep. A Customer Service Response form is included and customers are to complete if to acknowledge whether of not the facility is still in possession of an affected table. If so, the model number, serial number and facility name & address is to be completed so that Maquet reps can visit the facility and perform the corrective action.

Root Cause

Equipment maintenance

Distribution Pattern

Nationwide