Terminated

Volcano Imaging System s5i REF 807400001; CORE Mobile Imaging System (120V) REF 400-0100.01; CORE Mobile Imaging System Refurbished REF 400-0100.01-R; CORE Mobile Imaging System (240V) REF 400-0100.07; CORE Mobile Imaging System Refurbished REF 400-0100.07-R; CORE Mobile Imaging System (100V) REF 400-0100.08; CORE Mobile Imaging System Refurbished REF 400-0100.08-R; CORE Imaging System REF 400-0100.02 For intravascular ultrasound imaging.

FDA Recall Z-0189-2018 · Posted None

Recalling FirmVolcano Corporation
Recall StatusTerminated
Product CodeOBJ
Date Initiated2017-11-03
Date Terminated2023-10-20
Product Quantity1166 Units

Reason for Recall

On certain Impacted Systems, an unexpected Microsoft Windows Security dialog may appear during use, and the users response to the dialog may adversely affect the subsequent operation of the device.

Firm Action

Beginning the week of November 7, 2017, all affected US customers will be sent a field correction notification letter along with a Technical Service Bulletin (TSB) to explain how to assure the misconfiguration does not affect their system performance. For Distribution outside the US, the same letter and TSB will be sent to the initial consignee and they will execute the delivery of the Field Correction Notice and will inform the Recalling firm of the responses.

Root Cause

Software design

Distribution Pattern

Worldwide distribution including US nationwide and Canada, AE, AT, AU, BD, BE, BG, BR, CH, CL, CN, CY, DE, DK, EG, ES, FR, GB, HU, IE, IL, IN, IT, JO, JP, KW, KR, LB, MA, MX, MY, NL, NO, PA, PK, PL, PT, QA, RO, RU, SA, SE, SG, SK, Sri Lanka, TH, TW, UAE, UK, ZA,

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →