Monarch Airway Clearance System, Model 1000, Product Numbers: PMACS1NA (has blue tooth capability only) and PMACS1WI (has WIFI capability). The firm name on the label is Hill-Rom Services Pte. Ltd., Singapore. The Monarch Airway Clearance System, Model 1000, is intended to provide Airway Clearance Therapy and promote bronchial drainage where external manipulation of the thorax is the physicians choice of treatment. It is indicated for patients having difficulty with secretion clearance or the presence of atelectasis caused by mucus plugging.

FDA Recall Z-0192-2018 · Posted None

Recall StatusTerminated
Product CodeBYI
Date Initiated2017-09-15
Date Terminated2019-05-29
Product Quantity60 devices

Reason for Recall

Monarch Electromagnetic Interference (EMI) may disrupt the operation of implanted pacemakers or implantable cardioverter defibrillators as well as other devices that might be subject to an EMI if they come into direct contact with the Monarch device.

Firm Action

Phone calls were initiated September 15, 2017, to the customers, who are users. The expanded recall was initiated by letter dated November 13, 2017,via FedEx with delivery confirmation to both the users identified in the September notification and to the users who later received the recalled product. (Two local consignees will be provided the letter in person and one consignee will be sent the letter via USPS.) For further questions, please call (800) 426-4224.

Root Cause

Other

Distribution Pattern

USA (nationwide) Distribution to the states of: AZ, CA, CO, CT, FL, GA, IA, IL, IN, KY, MA, MI, MN, MO, NE, OH, OK, OR, TN, TX, VA, WA, WI, and WV. Internationally to Canada. There was no military/government distribution.