Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Baxter Clearlink Intravenous (IV) Delivery System; the system consists of solution sets, extension sets and blood sets that have a luer activated valve for IV access, which allows the administration of medications/solutions with a needleless luer syringe; Baxter Healthcare Corporation, Deerfield, IL 60015 USA

FDA Recall Z-0193-05 · Posted 2004-11-16

Recall StatusTerminated
Product CodeFPA
Date Initiated2004-04-28
Date Terminated2007-02-08
Product Quantity29,425,708 sets

Reason for Recall

Inadequate directions for use may have contributed to an increase in infection rates at a small number of customers beginning use of the Clearlink needleless device.

Firm Action

An Urgent Device Correction letter and reply form dated 4/28/04 were sent to all U.S. Clearlink customers, informing them of the reports of increased infection rates abd describing the revision of the product labeling with additional statements that relate to product usage and reinforcement of standard clinical practices and proper aseptic technique. The revised labeling statements are as follows: * Ensure clamp is open. Swab Luer activated surface with preferred antiseptic prior to first use and before every subsequent connection. * Don not swab Luer activated surface when clamp is closed or valve is recessed. Ineffective swabbibg may result. Replace if valve remains recessed. * Flush Luer activated valve after blood infusion/sampling. If valve cannot be cleared of blood, replace immediately. * Luer lock vacutainer holders, syringes and accessories are recommended for blood sampling. If blood remains on valve surface after sampling, swab surface. * If intermittently disconnecting set from Luer activated valve, immediately cover male Luer with a sterile replacement protector.

Root Cause

Other

Distribution Pattern

Nationwide, and internationally through Baxter subsidiaries to Macedonia, Nigeria, Saudi Arabia, South Africa, Ireland, Italy, Denmark, Germany, the Netherlands, Hong Kong, Indonesia, Singapore, Spain, Belgium, Norway, United Kingdom, France and Luxembourg.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑