Terminated

BIAD Classic 24 (24 inch collimator). The collimator is a component of the Trionix Biad Full Imaging System.

FDA Recall Z-0193-2007 · Posted 2006-11-16

Recall StatusTerminated
Product CodeKPS
Date Initiated2006-08-01
Date Terminated2008-10-29
Product Quantity76 devices

Reason for Recall

The ball screw on the unit which supports the detector head (which weighs approximately 800 lbs.) could snap resulting in the detector head dropping on the patient.

Firm Action

The recalling firm modified one of the devices in 2002. The remaining device users were telephoned on/about 11/02/2006. To date the recalling firm has not taken any corrective action regarding devices with serial numbers B-102, B-103, B-104 or B-106 thru B-184.

Root Cause

Other

Distribution Pattern

Worldwide, including USA, Australia, Egypt, UK, Belgium.