Axiom Artis Zee. Angiographic X-ray System.
FDA Recall Z-0194-2009 · Posted 2008-11-04
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeIZI
Date Initiated2008-08-22
Date Terminated2009-09-23
Product Quantity108 units
Reason for Recall
Patients could possibly fall off the table if not properly secured when: 1) moving them onto the patient table, 2) moving them on the patient table or 3) removing them from the patient table.
Firm Action
The recalling firm issued a Customer Safety Advisory Notice dated 7/04/08 (including an Addendum to the Operators Manual) which alerts the operator of possible injury to patients if they are not properly secured to the table during examinations and describes the recommended fastening procedures required to properly secure the patient to the table.
For questions or additional information, contact Siemens Medical Solutions, Inc. at 610-448-4634.
Root Cause
Labeling design
Distribution Pattern
Nationwide Distribution.