POLYSTAR, Angiographic X-Ray System, Model 1148902
FDA Recall Z-0195-05 · Posted 2004-11-16
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeIZI
Date Initiated2004-09-23
Date Terminated2005-01-25
Product Quantity43 units
Reason for Recall
Orbital gears may exhibit excess wear
Firm Action
The recalling firm issued Update Instructions, AX077/04/S, to Siemen''s Customer Service Representative (CSR) to visit affected sites and check the orbital gears for excessive wear. If the CSR finds gears exhibiting excessive wear the orbital gears will be replaced. If no wear is observed the gears will be lubricated as a preventive measure.
Root Cause
Other
Distribution Pattern
The product was shipped to medical facilities nationwide and government accounts in KS and MI.