Analogic SyneRad Multi Digital Radiology System, Model AN6150
FDA Recall Z-0196-06 · Posted 2005-11-24
Advertisement
Recalling FirmAnalogic Corporation
Recall StatusTerminated
Product CodeKPR
Date Initiated2005-10-14
Date Terminated2015-04-14
Product Quantity17 UNITS
Reason for Recall
Potential weld defect in the Overhead Tube Support assembly which connects to the telescoping arm. This could cause the tube assembly to fall, leading to possiible injury of the user or patient.
Firm Action
Analogic Corp notified accounts by email on 10/14/2005 and by letter on 10/17/2005. Accounts are requested to stop use and correction will be conducted by a service engineer.
Root Cause
Other
Distribution Pattern
AZ, CA, FL, MA, OH, PA, TN, TX,