Optetrak, B-series cemented finned tibial tray size 2f/2t. Catalog # 220-04-02
FDA Recall Z-0196-2007 · Posted 2006-11-21
Advertisement
Recalling FirmExactech, Inc.
Recall StatusTerminated
Product CodeJWH
Date Initiated2006-08-31
Date Terminated2012-05-03
Product Quantity30
Reason for Recall
Improperly machined parts: The Cemented Finned Tibial Trays could not seat the size 2 mating trial or polyethylene insert.
Firm Action
The recall was initiated on August 31, 2006. The distributors were notified by FAX, email and telephone. The returned devices will be stored quarantine.
Root Cause
Other
Distribution Pattern
International distribution only---- including countries of South America, Africa and Europe