Terminated

MR Surgical Suite II Table, Model Number(s): M8074SS, M8053SS, M0074SS. UPDATE 11-4-2015: Model M0001SS

FDA Recall Z-0196-2016 · Posted 2015-10-30

Recall StatusTerminated
Product CodeLNH
Date Initiated2015-09-23
Date Terminated2016-03-28
Product Quantity23 (11 US; 12 OUS) Update 11-4-2015: 35 (14 US; 21 OUS)

Reason for Recall

GE became aware of an issue with the table transfer release mechanism of the MR Surgical Suite Option. The issue may prevent patient transfer between the MR and surgical tables.

Firm Action

GE sent an "Urgent Medical Device Correction" letter dated September 23, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. You may continue to use the table. Monitor for any issues in releasing the table. If you see this issue, contact a GE Field Engineer. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution - US Distribution including the states of : CA, CO, FL, IL, MI, OH, TX and WI., and to the countries of : AUSTRIA, BRAZIL, FRANCE, GERMANY, JAPAN, POLAND and TURKEY. UPDATE 11-4-2015: Additional states: AZ, MN. Additional countries: Canada, China, Czech Republic, South Africa, and United Kingdom

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.