Terminated

DEMO No Needle Corrugated Carton 30 U/D, Part Number 2260-02

FDA Recall Z-0196-2020 · Posted None

Recall StatusTerminated
Product CodeLZG
Date Initiated2019-07-03
Date Terminated2020-06-24
Product Quantity20,532 devices

Reason for Recall

There is a potential for "No Needle Demo Units" to contain a needle.

Firm Action

On July 3, 2019, Valeritas issued notices to customers via first class U.S. postal service. The removal strategy is to notify active HCP office accounts from the manufacturing start date of effected product (January 2018) who may have No Needle Demo units in their possession. The removal letter instructs HCP offices to dispose of any No Needle Demo product on hand and acknowledgement by signing and returning the removal letter as instructed.

Root Cause

Employee error

Distribution Pattern

Nationwide within U.S.