Cytomics FC 500 Flow Cytometry System with CXP Software Versions 2.0 & 2.1
FDA Recall Z-0197-2007 · Posted 2006-11-21
Advertisement
Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeGKZ
Date Initiated2006-07-26
Date Terminated2012-02-29
Product Quantity416 (US) + 60 (Canada)
Reason for Recall
Mis-identification-If a panel or protocol is added to an existing worklist but the tube location is not specified, the CXP Acquisition software will run the last specified tube through the remaining protocols and will generate results and printouts of these runs leading to a Mis-Identification condition.
Firm Action
Customer letter -- A Product Corrective Action (PCA) letter was sent the week of July 24, 2006 to all accounts that have FC500 with CXP Software 2.0 and 2.1 to inform them that a potential sample misidentification may occur when running a Work list with missing tube Location(s) in Automatic MCL mode. The letter outlines actions to be taken to avoid mis-identification of patient results.
Root Cause
Other
Distribution Pattern
Nationwide and Canada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.