GE Healthcare Optima XR 200amx with digital upgrade Mobile general purpose radiographic imaging of the human head and body.
FDA Recall Z-0197-2013 · Posted 2012-11-16
Advertisement
Recalling FirmGE Healthcare, LLC
Recall StatusTerminated
Product CodeOXO
Date Initiated2012-09-20
Date Terminated2013-12-05
Product Quantity364 total units installed in US
Reason for Recall
It was discovered GE Healthcare has become aware of a firmware issue on the mobile radiographic product, Brivo XR285amx, Optima XR200amx, Optima XR22Oamx and Optima XR200amx with Digital Upgrade, related to Technique Accuracy as mandated by the US Code of Federal Regulations (21CFR) and IEC 60601-2-54.
Firm Action
GE Healthcare will remedy the defect or bring the product into compliance with each applicable Federal and IEC standard. A Service Representative will update the firmware on the system to address the issue.
Field Modification Instructions (FMI) 10870, 10872, and 10873 are describing the rework plan the GEHC Engineers will make to the imaging systems compliant with the applicable performance standard.
Further questions please call (262) 513-4122.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
Nationwide Distribution
Official Source