BD PhoenixTM AP, Catalog Number 448010 The BD Phoenix" AP instrument is designed for use with the BD Phoenix" system is intended to standardize ID broth tube inoculum, add AST indicator to the AST broth tube and transfer an aliquot of ID broth to AST broth tubes, as required for preparing samples for use on the BD Phoenix" system, which performs identification and susceptibility testing
FDA Recall Z-0198-2018 · Posted None
Recalling FirmBecton Dickinson & Co.
Recall StatusTerminated
Product CodeJQX
Date Initiated2017-06-01
Date Terminated2018-05-09
Product Quantity7 units
Reason for Recall
Potential unexpected movement of robot arm
Firm Action
BD Diagnostic Systems telephoned all customers on June 1, 2017 and sent an Urgent Product Recall letter dated May 31, 2017, and Response Forms to their customers via email. BD recently confirmed a manufacturing issue with the motor that drives movement of the robot arm. The issue may result in the inability of this motor to withstand the lateral forces encountered during normal operations. Should this occur, the instrument will indicate an E20 error-Pipettor step loss during operation and stop functioning. While there is no loss of patient sample, the samples would need to be manually processed until the instrument can be repaired. To limit any disruption in the workload, BD Service will contact customers to schedule the instrument replacement.
Customers are advised to complete and return the response form via email to [email protected]. Customers may continue to use the instrument until replacement is performed. Customers with questions may contact BD (Customer/Technical) Support at 1-800-638-8663.
For questions regarding this recall call 410-316-4258.
Root Cause
Process control
Distribution Pattern
Worldwide Distribution - US including CA, NC & TN Internationally: Canada
Official Source