Recall StatusTerminated
Date Initiated2013-10-28
Date Terminated2014-12-24
Product Quantity18 units
Reason for Recall
Toshiba America Medical Systems, Inc. is recalling due to a potential problem with Specific Absorption Rate (SAR)
values when using the Toshiba Titan 3T System.
When the system is used in combination with either the QD Head Coil (MJQH-142A) or the Large Knee
SPEEDER (MJAJ-182A), SAR values lower than the actual SAR values may be displayed on the operating
monitor.
Firm Action
Toshiba America Medical Systems, Inc. sent an "URGENT: MEDICAL DEVICE CORRECTION". letter dated October 28, 2013 via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS for retention. The letter listed Device, Serial Number, reason for recall, problem, corrective actions, request to customers, and contact information.
Root Cause
Nonconforming Material/Component
Distribution Pattern
USA Nationwide Distribution in the states of: CA, FL, GA, IA, NV, NY, LA, IN, PA, OR, TX, and WI.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.