MaxAn 4.0mm and 4.5mm fixed screws, used in the MaxAn Anterior Cervical Plate System. The MaxAn Anterior Cervical Plate System is a decompression-based technique for cervical spine stabilization.
FDA Recall Z-0203-2016 · Posted 2015-11-03
Advertisement
Recalling FirmBiomet Spine, LLC
Recall StatusTerminated
Product CodeKWQ
Date Initiated2015-09-29
Date Terminated2016-04-29
Product Quantity17 units
Reason for Recall
The packaging of 4.5 mm diameter MaxAn screws identifies/labels the screws as 4.0 mm diameter screws.
Firm Action
Zimmer Biomet Spine sent an Urgent Medical Device Recall Notice dated September 29, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
If the device has already been implanted, written confirmation of this will be required and the necessary parties will be contacted. They will store returned product in quarantine in an isolated location until they can undergo inspection to verify if they are conforming or not and then will be destroyed or retained. Necessary parts will be destroyed so that it is visually obvious they are no longer functional. For further questions call (303) 443-7500 ext.244.
Root Cause
Labeling mix-ups
Distribution Pattern
US Distribution to the states of : NY, MO, and FL
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement