McKesson PREVENT HT Safety Blood Collection Needles, 21G 1 1/4 inch, MFR # 16-NBC21GS

FDA Recall Z-0204-2022 · Posted None

Recall StatusTerminated
Product CodeFMI
Date Initiated2021-08-24
Date Terminated2023-07-21

Reason for Recall

Potential for difficulty in safety mechanism activation.

Firm Action

On 8/24/2021, Zhejiang Kindly Medical Devices Co., Ltd. issued Urgent Medical Device Safety Alert notices to two customers. On 9/17/2021, issued a second notice to customers dated 8/24/2021. Action to be Taken by User: 1 - Advise the user to lower the angle between safety protective cap and needle tube by slight pressing in the direction of activation, then using a thumb or hard surface to activate the safety protective cap and lock the needle tube in the IFU, then together with corresponding illustration. 2 - Request the user to strictly follow the guidance of the IFU.

Root Cause

Nonconforming Material/Component

Distribution Pattern

U.S. Nationwide distribution.