Terumo Advanced Perfusion System 1 base, 110/120V; Model 801763. For use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion and cardiopulmonary bypass procedures.

FDA Recall Z-0205-2009 · Posted 2008-10-31

Recall StatusTerminated
Product CodeDTQ
Date Initiated2008-07-18
Date Terminated2009-07-07
Product Quantity626

Reason for Recall

Local pump controls including start/stop button, select button and manual button may not function.

Firm Action

Consignees were notified by recall letter dated 7/18/08 and were provided with revised operating instructions telling them to revert to the central control module should these problems occur.

Root Cause

Component design/selection

Distribution Pattern

Nationwide, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arab Emirates and Viet Nam.