AD5 table for use with Integris and Integris Allura x-ray systems
FDA Recall Z-0206-06 · Posted 2005-11-29
Advertisement
Recalling FirmPhilips Medical Systems North America Co. Phillips
Recall StatusTerminated
Product CodeIZI
Date Initiated2005-10-12
Date Terminated2011-03-09
Product Quantity1770 units with AD5 tables
Reason for Recall
Potential for high voltage exposure involving the injector interface connector of patient support tables.
Firm Action
On 10/10/05 the firm issued a letter to customers alerting customers to the problem and advising that the firm will perform an upgrade to the devices at customer sites. In March 2006 the firm issued a second notification letter involving more units.
Root Cause
Other
Distribution Pattern
Devices were distributed nationwide to hospitals/medical centers.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K031333PHILIPS INTEGRIS ALLURA FLAT DETECTOR RELEASE 1.2Philips Medical Systems North America, Inc. · 2003-05-13K020055PHILIPS INTEGRIS ALLURA 9 WITH FD OPTIONPhilips Medical Systems · 2002-03-15K002016PHILIPS INTEGRIS ALLURAPhilips Medical Systems North America, Inc. · 2000-09-06K984545RELEASE 2 OF PHILIPS INTEGRIS SERIES (V5000, BV5000, H5000, BH5000, HM2000) SYSTEMSPhilips Medical Systems North America, Inc. · 1999-02-25K925302PHILIPS INTEGRIS H/HM SYSTEMPhilips Medical Systems, Inc. · 1992-12-17