LifeStar ACT I Sensor Ambulatory Cardiac Telemetry; an automatically activated system with a single channel EKG, office hookup and auto send/auto detect that requires no patient intervention to either capture or transmit an arrhythmia when it occurs; Manufactured by Card Guard, 2 Pekeris St, Rehovot, 76100, Israel; Model FG-00065 LifeStar ACT provides critical ECG information by capturing the onset and termination/escape of an arrhythmia to assist in correctly identifying and treating the patient. Patients are monitored for up to 30 days and all of the data is available at the request of the physician during the service period and for 7 days following the end of service.

FDA Recall Z-0206-2012 · Posted 2011-11-17

Recall StatusTerminated
Product CodeDXH
Date Initiated2011-09-26
Date Terminated2012-05-15
Product Quantity120 sensors

Reason for Recall

LifeWatch received complaints of minor shocks and burns during use of the LifeStar ACT systems with shorts in the leads.

Firm Action

The firm, LifeWatch, sent an " Urgent Device Recall" letter dated September 27, 2011 to the affected customers. The letter describes the product, problem and action to be taken. The customers were instructed to immediately refrain from providing these devices to their customers; stop using the affected sensors, return their entire kit in the pre-paid mailer, and use the new device kit included with the recall letter. In addition, the letter informed the users that the replacement device has upgraded software and firmware that allows for detection of shorts in the cable, and if detected, cuts off current flow to the patient to prevent minor shocks and burns. If you have any questions, please call (847) 813-4258.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide distribution.