Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Diasol Acid Concentrate REF G100325-10Dex 100

FDA Recall Z-0207-2025 · Posted 2024-10-23

Recalling FirmDiasol, Inc
Recall StatusOpen, Classified
Product CodeKPO
Date Initiated2022-07-27
Product Quantity852 gallon bottles

Reason for Recall

Due to labeling issue. The box label is different that the gallon label in that gallon label shows a lower Calcium concentration than the box label.

Firm Action

On 07/26/2022, the firm communicated via email with customer that 2 case were incorrectly labeled. Cases marked 3K 2.5ca have inside box label as 3K 2.0ca. Customer was instructed not to use the product and the firm picked up mislabeled product.

Root Cause

Labeling mix-ups

Distribution Pattern

U.S. Nationwide distribution in the state of AZ.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑