Terminated

Sechrist Monoplace Hyperbaric Chamber with Electronic Control Module Model 3600E/ER The Hyperbaric Chamber is intended to provide 100% oxygen to the patient at greater than ambient pressures

FDA Recall Z-0209-2013 · Posted 2012-11-06

Recall StatusTerminated
Product CodeCBF
Date Initiated2008-08-28
Date Terminated2012-11-06
Product Quantity65 units total (61 units in the US)

Reason for Recall

Sechrist conducted a field upgrade on the hyperbaulic chambers equipped with Electronic Control Modules in order to correct a software bug that could potentially affect the performance of the chamber.

Firm Action

Sechrist Industries contacted consignees by phone to notify them of a software upgarde and to schedule appointments to perform the required field upgrades. An approximate summary of the phone notification is provided below. "I am contacting you to notify you of a field upgrade we are performing on our hyperbaric chambers equipped with Electronic Control Modules in order to correct a software bug that could potentially affect the performance of your chamber. The upgrade includes the installation of a new motherboard with revised software, a reset switch, and the exchange of other control panel components. The process will take approximately 2.5 hours and will be performed at no charge to you." For questions regarding this recall call 714-579-8400.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution - USA (nationwide) and the countries of Bulgaria, and Canada