Beckman Coulter Lipase, REF OSR6130 and OSR6230.
FDA Recall Z-0210-2020 · Posted None
Advertisement
Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeCDT
Date Initiated2018-08-24
Date Terminated2024-08-02
Product Quantity20,926 units
Reason for Recall
NAPQI, a metabolite of Acetaminophen, causes negative interference with the recalled product assays at levels present in the serum/plasma of overdose patients.
Firm Action
The recalling firm issued letters dated 8/21/2018 via email or postal mail if an email address was not available on 8/24/2018 to their customers. The letter identified the affected product, problem and actions to be taken. For questions call 1-800-223-0139 in the United States and Canada. Outside the United States and Canada contact local Beckman Coulter Representative.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Distribution was nationwide. There was government and military distribution. Foreign distribution was made to Canada.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.