Covidien Emprint Ablation Catheter with Thermosphere Technology Bronchoscopic Ablation, CFN CA108L1
FDA Recall Z-0210-2022 · Posted None
Advertisement
Recalling FirmCovidien Llc
Recall StatusTerminated
Product CodeNEY
Date Initiated2021-09-29
Date Terminated2025-02-13
Product Quantity14 units
Reason for Recall
The firm identified an error in the device temperature monitoring circuitry which is designed to interrupt energy delivery in the event of a primary cooling system failure. Use of an affected device in conjunction with a primary cooling system failure could lead to the applicator tip overheating and associated heat damage which could result in an allergic/toxic reaction, unintended tissue ablation, tissue burn, and/or a delay of treatment while an alternate device is located.
Firm Action
Beginning September 29, 2021, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were asked to immediately quarantine and discontinue use of the affected products. Customers should return affected product to the firm. If product has been further distributed, please forward the recall information. Customers should complete and return the accompanying response form even if they do not have product on hand.
Root Cause
Process control
Distribution Pattern
No domestic distribution. Foreign distribution to Hong Kong and UK.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.