MotoBand CP, Poly-axial Locking Screw, 3.0mm x 28mm

FDA Recall Z-0212-2020 · Posted None

Recall StatusTerminated
Product CodeHWC
Date Initiated2019-09-19
Date Terminated2021-12-02
Product Quantity36 screws

Reason for Recall

Potentially includes a 3.0mm x 30mm screw instead of 3.0mm x 28mm screw.

Firm Action

CrossRoads Extremity Systems notified customers on about 09/19/2019 via email. The email informed customers that pack-out potentially includes a 3.0mm x 30mm screw instead of 3.0 x 28mm screw. Instructions included to return all affected screws to CrossRoads Extremity Solutions immediately with provided label and return materials authorization.

Root Cause

Labeling mix-ups

Distribution Pattern

Nationwide distribution to AK, CA, FL, GA, IL, MA, MN, NC, NJ, OK, OR, WA.